K-TC: calibrating channels and updating firmware
Bring each K-TC channel into line with a calibrated reference, clear a calibration you no longer want and load new firmware safely.
Before you start
- K-CAL connected to the module. See K-TC: configuring channels and CAN output in K-CAL.
- For calibration, calibrated thermocouple test equipment that can apply each reference temperature to an input.
- For a calibration certificate, ask through the contact page. Influx Technology offers calibration and certification on demand.
Calibration tab
The Calibration tab steps you through a set of reference points per channel. When you commit, K-CAL writes the new calibration to the module.

- Load
- Reads the calibration stored in the module. Use it to check whether the module is calibrated.
- RUN
- Calibrates the selected point on the selected channel.
- Reference
- The temperature you apply to the channel for this point.
- Measured Temperature
- The raw temperature the channel reads now.
- Noise
- The noise level on the signal.
- Settings
- Sets the reference temperatures. Go to next channel moves on to the next channel after each point.
- Noise Level
- The noise tolerances. Tolerance Default restores the defaults.
- Clear calibration
- Clears the whole calibration stored in the module.
- Clear channel
- Clears the calibration of one channel.
Calibrate the channels
- Choose the reference points. Use Settings to set reference temperatures that cover the range you expect to measure.
- Select the channel and point. Use the channel's radio button, for example TC 0, or select the point in the table.
- Apply the reference. Connect your test equipment to the channel and set it to the reference temperature shown in Reference.
- Select RUN. If the input is stable and close to the reference, the point is updated in the table.
- Repeat. Work through every point on every channel you want to calibrate.
- Write the calibration. Select Commit. K-CAL saves the calibration to the module and creates a calibration report.
If the measured value is too far from the reference, K-CAL warns you before it calibrates the point.

The report lists the module's serial number and firmware version, the date and time, the noise tolerances and the pass or fail result for each channel. Keep it with your test records.
Watch: calibrate the thermocouple channels (1:39)
Clear the calibration
- Start the clear. On the Calibration tab, select Clear calibration. To clear only one channel, select it, then select Clear channel instead.
- Confirm. Select Yes.
- Watch the status messages. They appear at the bottom of the K-CAL window. If you cannot see them, drag the light blue line at the bottom of the window upwards.
- Wait for the confirmation. The status messages show when the calibration table has been cleared.
Update the firmware
- Open the Reprogramming tab. The toolbar's first button becomes Open.
-
Select the firmware file. Select Open, then select the
.hexfile. Firmware files are installed with K-CAL in itsfirmwarefolder, for exampleC:\Program Files (x86)\Influx Technology\Influx K-Cal\firmware. - Check the file. The tab shows the file's name, dates, size and its boot and flash software versions.
- Start the update. Select Commit. A progress bar runs along the bottom of the window.
- Wait for it to finish. The firmware update is complete when the progress bar reaches the right-hand side.