K-BOX: calibrating inputs and updating firmware
Calibrate the K-BOX inputs against a calibrated source to improve accuracy, clear a calibration that went wrong and update the module firmware.
Before you start
- K-BOX connected to K-CAL. See K-BOX: configuring analogue, thermocouple and digital channels.
- A calibrated voltage source or thermocouple simulator for each point you want to calibrate.
- For a calibration certificate, contact us. Influx Technology offers calibration and certification on demand.
The Calibration tab
Open the Calibration tab and select Load to read the calibration stored in K-BOX.

- RUN
- Calibrates the selected point of the selected channel.
- Reference
- The value of the calibration point, a voltage for an analogue channel or a temperature for a thermocouple.
- Measured value
- What K-BOX reads on the selected channel now.
- Noise
- A bar that shows how noisy the input is.
- Settings
- With Go to next channel ticked, K-CAL moves to the next channel after each point.
- Noise Level
- The two noise thresholds the Noise bar uses. Tolerance Default restores them.
- Analog Inputs and Thermo Couples
- Buttons to select the channel, ADC 0 to ADC 7 and TC 0 to TC 7.
- Clear calibration and Clear channel
- Remove the calibration of the module, or of the selected channel.
Calibrate an input
Choose calibration points that suit the signals you expect. For example, if a channel will see 0 V to +38 V, calibrate its 0 V and +40 V points.
- Connect the reference source. Wire the calibrated source to the input you are calibrating.
- Choose the point. Select its value in the Reference list.
- Select the channel. Use its button under Analog Inputs or Thermo Couples, or select the point in the grid.
- Set the source. Adjust the calibrated source until it matches the reference value.
- Select RUN. If the input is stable and correct, K-CAL updates the calibration of that point.
- Repeat for every point. Do the same on each channel you need.
- Select Commit. K-CAL saves the calibration to K-BOX and creates a report of the calibration.
If the reading and the reference are too far apart, K-CAL warns you before it calibrates the point. Unless you are sure the reference is right, select No and check the wiring and the source.

Clear the calibration
If a calibration went wrong, clear it and start again.
- Select Clear calibration. To clear only the selected channel, select Clear channel instead.
- Confirm. Select Yes.
- Watch the status messages. They appear at the bottom of the K-CAL window. If you cannot see them, drag the light blue divider upwards.
- Wait for the confirmation. The messages end with "Calibration table was cleared."

Update the firmware
K-CAL installs the K-BOX firmware as a .hex file, usually in C:\Program Files (x86)\Influx Technology\K-Box Config\firmware. For the latest K-CAL, see Software downloads: installers, drivers and firmware.
- Open the Reprogramming tab.
-
Select Open. Find the hex file, for example
KBox_B05_F032.hex, and select Open. - Select Commit. A progress bar appears at the bottom of the window.
- Wait for the bar to fill. The firmware update is complete when the bar reaches the right-hand side.