K-BOXSpecification

K-BOX: overview and specifications

Updated

Find out what K-BOX measures, how it reaches your logger and the full specification, so you can plan the channels, power and wiring before you install it.

This is step 1 of the six-part K-BOX series. The series covers:

  • The specification
  • The connector pinouts
  • Setting up channels in K-CAL
  • The digital measurement modes
  • Conversion tables and formulas
  • Calibration and firmware

What K-BOX does

K-BOX (INF2201) is the multi-DAQ module of the K-Series. It measures thermocouples, voltages, currents and PWM signals at the same time and sends the results periodically on the CAN bus. The measurement data is time synchronised with the vehicle network data your logger records.

K-BOX has no storage of its own. Any CAN logger records it, such as a REBEL, a REXGEN or a third-party logger. A DBC file describes the messages, so any tool that reads DBC files decodes them. Set the CAN IDs, rates, ranges and calibration in K-CAL. K-BOX keeps them when the power is off.

  • Stack it with other K-Series modules. The two CAN connectors carry CAN and power from one module to the next.
  • Use it for applications with many thermocouples and analogue inputs.
  • Leave it on a vehicle. WakeOnCAN and deep sleep suit unattended fleet installs.
K-CAL with the General tab open, showing baud rate, byte order, sleep mode, tool options and the list of CAN interfaces
You configure K-BOX over CAN in K-CAL. The title bar of this version reads K-Box Cal.

How K-BOX connects to a logger

K-BOX connects to the outside world only through CAN. It has no PC port, so you configure it in K-CAL through a CAN interface or an Influx Technology logger.

  • With a REBEL logger: the REBEL can power the module through the multi-connect cable when its auxiliary power output is enabled.
  • With a REXGEN logger: give the module an external supply. A REXGEN does not provide power for a K-Series module.
  • In a stack: give every extra module an external supply.

For each case in detail, see these guides:

Key features

  • Eight bipolar differential analogue inputs on software switchable ranges of ±80 V, ±40 V, ±20 V or ±10 V.
  • Up to eight type J, K or T thermocouples with cold junction compensation.
  • Four digital channels, three of which measure frequency, pulses, duty cycle, events or RPM up to 100 kHz.
  • Regulated 5 V and 24 V outputs to power external sensors.
  • Current measurement on the analogue inputs through a current clamp or shunt.
  • Galvanic isolation between the enclosure, power, CAN bus, analogue board and thermocouple board.
  • Configuration through a DBC file and K-CAL for Windows.

Specifications

These tables give the full K-BOX specification.

Power and interfaces

K-BOX power and interfaces
Item Specification
Supply voltage 6 to 36 V DC
Power consumption Typically under 3 W
Power saving WakeOnCAN and power-down deep sleep
Sensor supply outputs Regulated 5 V and 24 V, 80 mA combined
CAN interface CAN 2.0, data transmitted periodically and time synchronised with the vehicle network data
PC interface None. Configure over CAN in K-CAL through a CAN interface or an Influx Technology logger
Signal definition DBC file, exported from K-CAL
Settings Stored in the module and kept without power
Isolation Galvanic isolation between enclosure, power, CAN bus, analogue board and thermocouple board
Indicators Two LEDs, Status and CAN

Analogue inputs

K-BOX analogue inputs
Item Specification
Channels 8 bipolar differential inputs
Ranges Software switchable ±80 V, ±40 V, ±20 V or ±10 V
Resolution 16 bit
Accuracy ±0.0015% of full scale
Sampling rate 1 kHz on all 8 channels, 2 kHz on 4 or 10 kHz on 2
Input impedance Above 4 MΩ
Over-voltage protection ±150 V on the analogue channels
Current Measured on the analogue inputs through a current clamp or shunt

Thermocouple inputs

K-BOX thermocouple inputs
Item Specification
Channels Up to 8, type J, K or T
Compensation Automatic cold junction compensation
Range −200 °C to +1250 °C
Accuracy ±1 °C
Resolution 0.1 °C
Sampling rate 10 Hz on all 8 channels
Maximum input voltage ±3.3 V
Connectors Miniature flat-pin thermocouple sockets

Digital inputs and outputs

K-BOX digital channels
Item Specification
Channels 4 unipolar single-ended channels, each set up as an input or an output
Measurement inputs Digital 1 to 3, for pulse duration, frequency, duty cycle, event count, cycle duration or RPM, up to 100 kHz
Pulse timing Minimum pulse 100 ns, minimum 10 µs between pulses
Input thresholds Low below 1.5 V, high above 2.0 V, up to 12 V
Input voltage range −8 V to +12 V
Input leakage current Under 10 nA
Outputs Open collector outputs, 45 V and 10 mA maximum
Saturation voltage, output on Under 0.15 V
Equivalent on-resistance Under 510 Ω
Leakage current, output off Under 5 µA

Enclosure and environment

K-BOX enclosure and environment
Item Specification
Enclosure Stackable, daisy-chainable ABS
Ingress protection IP65
Dimensions 115 × 105 × 79 mm
Weight 450 g
Operating temperature −40 °C to +85 °C
Humidity Up to 90%
Connectors 25-way D-Sub for the analogue inputs, two 9-way D-Sub for CAN in and out, 8 thermocouple sockets
Part number INF2201, sold as the INF2201P package

Download the K-BOX spec sheet (PDF, 53 KB)

Warranty, calibration and support

K-BOX carries a 12-month warranty. See the returns and warranty policy. Influx Technology offers calibration and certification on demand. You can also calibrate the inputs yourself in K-CAL, in step 6 of this series. For help with a module, contact technical support.

Next: K-BOX: connector pinouts.