K-BOXPinout
K-BOX: connector pinouts
Use these tables to wire power, CAN, sensors and thermocouples to K-BOX, and to stay within the limits of each pin.
Connectors at a glance
K-BOX (INF2201) has these connectors and indicators:
- Two 9-way D-Sub CAN connectors with nuts, one male and one female. Both carry the same signals, so CAN and power pass through a stack of K-Series modules.
- One 25-way male D-Sub, labelled ADC, for the eight analogue inputs and the sensor supplies.
- Eight miniature flat-pin thermocouple sockets, labelled TC0 to TC7.
- Two LEDs, labelled Status and CAN.
Pins missing from the tables are not connected.
CAN connectors
| Pin | Function |
|---|---|
| 1 | Digital 4 output, or +4.5 V instrumentation supply |
| 2 | CAN-L |
| 3 | Ground |
| 4 | Digital 1, input or output |
| 5 | Power ground |
| 6 | Digital 3, input or output |
| 7 | CAN-H |
| 8 | Digital 2, input or output |
| 9 | Supply voltage, 6 to 36 V DC |
Digital pin limits
- As inputs: keep the voltage between −8 V and +12 V. An input reads low below 1.5 V and high above 2.0 V.
- As outputs: the open collector outputs are rated 45 V and 10 mA. Drive larger loads through a relay.
Only Digital 1 to 3 can measure signals. For the input and output modes, see K-BOX: configuring analogue, thermocouple and digital channels.
ADC connector
The ADC connector is a 25-way male D-Sub with nuts. Each analogue input is a differential pair. For example, input 0 uses pins 6 and 18, and input 7 uses pins 13 and 25.
| Pin | Function |
|---|---|
| 1 | +24 V sensor supply output |
| 2 | +5 V sensor supply output |
| 5 | Analogue ground |
| 6 | Analogue input 0 positive |
| 7 | Analogue input 1 positive |
| 8 | Analogue input 2 positive |
| 9 | Analogue input 3 positive |
| 10 | Analogue input 4 positive |
| 11 | Analogue input 5 positive |
| 12 | Analogue input 6 positive |
| 13 | Analogue input 7 positive |
| 14 | Ground out |
| 15 | Ground out |
| 17 | Analogue ground |
| 18 | Analogue input 0 negative |
| 19 | Analogue input 1 negative |
| 20 | Analogue input 2 negative |
| 21 | Analogue input 3 negative |
| 22 | Analogue input 4 negative |
| 23 | Analogue input 5 negative |
| 24 | Analogue input 6 negative |
| 25 | Analogue input 7 negative |
- Sensor supplies: the +24 V and +5 V outputs share 80 mA between them.
- Analogue inputs: do not apply voltages outside the range you set for the channel in K-CAL. The channels are protected to ±150 V.
Thermocouple sockets
The eight thermocouple sockets are miniature flat-pin sockets. Plug a miniature flat-pin thermocouple plug of type J, K or T into each one. The sockets fitted are green, the IEC colour for type K.
| Contact | Function |
|---|---|
| Top, narrow blade | Thermocouple positive |
| Bottom, wide blade | Thermocouple negative |
| Dimension | mm | inches |
|---|---|---|
| Blade spacing, centre to centre | 8 | 0.31 |
| Positive blade width | 2.4 | 0.09 |
| Negative blade width | 3.17 | 0.13 |
| Blade length | 12.7 | 0.5 |
| Body length | 29.57 | 1.2 |
| Body width | 20.6 | 0.81 |
| Cable clamp | 40 AWG wire up to 5.5 mm cable | up to 0.22 |
Wiring the digital pins
A switch, a transistor, a relay contact or a logic buffer can drive a digital input. Each pulls the input low against a pull-up resistor.
An open collector output switches a load to ground. Supply the load from an external source and fit a flyback diode across a relay coil.
For the full set of input and output circuits, see REBEL and K-Series: wiring the wake-up pin and digital I/O. For the cables that link modules, see K-Series link cables, termination and terminals.
Next: K-BOX: configuring analogue, thermocouple and digital channels.